JUM AIUM Career Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Order Full text via Infotrieve
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nelson, T. R.
Right arrow Articles by Pretorius, D. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nelson, T. R.
Right arrow Articles by Pretorius, D. H.

Journal of Ultrasound in Medicine, Vol 9, Issue 10 575-581, Copyright © 1990 by American Institute of Ultrasound in Medicine


JOURNAL ARTICLE

Device for the calibration of flow-velocity-measuring Doppler ultrasound equipment

T. R. Nelson and D. H. Pretorius
Department of Radiology, University of California, San Diego, La Jolla 92093.

A new device is described for verifying the calibration accuracy of flow-velocity Doppler ultrasound (US) equipment. A rotating circular disk whose circumferential velocity may be fixed or modulated provides a strong signal source suitable for calibrating clinical Doppler US instruments. Circumferential edge velocity, derived from the disk angular frequency, is used to verify the accuracy of Doppler instrument calibration. A limited survey of routinely used Doppler US instruments demonstrates that calibration accuracy generally is satisfactory for clinical purposes, with minimal variation between instruments. Factors that affect measurements include transducer type, scale range, and display size. All transducers used on a specific piece of equipment need to be tested because the transducer design is an important part of the instrument operation. Variations were observed in reported velocities obtained from different transducers on the same equipment and for the same transducer with different scale ranges. In particular, some scales were more accurate than others, suggesting that individual scales should be tested when performing clinical Doppler US instrument calibration. Other measures of performance including scale display range, sample depth, aliasing, and transducer selection may be studied to identify their effect on measurement accuracy. Temporal modulation of angular velocity produces waveforms that may be used to simulate pulsatile velocity characteristics. The device is simple to construct and use, and it provides a convenient way to verify high-quality instrument performance.


This article has been cited by other articles:


Home page
StrokeHome page
O. E. H. Elgersma, M. van Leersum, P. C. Buijs, M. S. van Leeuwen, Y. T. van de Schouw, B. C. Eikelboom, and Y. van der Graaf
Changes Over Time in Optimal Duplex Threshold for the Identification of Patients Eligible for Carotid Endarterectomy
Stroke, November 1, 1998; 29(11): 2352 - 2356.