JUM Parker Laboratories, Inc
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pemberton, J.
Right arrow Articles by Sahn, D. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pemberton, J.
Right arrow Articles by Sahn, D. J.
© 2005 by the American Institute of Ultrasound in Medicine
J Ultrasound Med 24:1109-1115 • 0278-4297

Accuracy of 3-Dimensional Color Doppler-Derived Flow Volumes With Increasing Image Depth

James Pemberton, MD, Ling Hui, MD, PhD, Monica Young, Xiaokui Li, MD, Antoinette Kenny, MD and David J. Sahn, MD

Oregon Health & Science University, Portland, Oregon, USA (J.P., L.H., M.Y., X.L., D.J.S.); and Department of Cardiology, The Freeman Hospital, Newcastle Upon Tyne, England (A.K.).

Address correspondence to David J. Sahn, MD, L608, Pediatric Cardiology, Oregon Health & Science University, 3181 SW Sam Jackson Park Rd, Portland, OR 97239-3098 USA. E-mail: sahnd{at}ohsu.edu

Objectives. We and others have reported on the use of digital color Doppler sonography from real-time 3-dimensional (3D) echocardiography and its use in accurately calculating cardiac flow volumes, namely stroke volume (SV) and, hence, cardiac output. However, in some patients, image depth is higher than average, and this may affect the accuracy of volume calculation. We sought to investigate the impact of image depth and the accompanying change in signal strength, spatial resolution, and pulse repetition frequency on the accuracy of SV calculation from 3D color Doppler data in an in vitro model. Methods. A tube model of the left ventricular outflow tract was constructed from plastic tubing and connected to a pulsatile pump. The volume flowing through the tube was imaged using a 3D echocardiography system. Stroke volumes from the pump were computed from the DICOM data using commercially available software and compared with a reference standard of timed volumes with the use of a graduated measuring cylinder over a range of depth settings and SVs. Results. There was good correlation between the 3D-derived SVs and the reference cylinder measures over all depths from 4 to 16 cm at 1-cm increments with a tube diameter of 17 mm, a pump rate of 60 beats/min, and SVs ranging from 20 to 70 mL. The average r2 value for the 13 different depths was 0.976. However, the accuracy of the 3D method of volume calculation appeared to fall at depths greater than 13 cm, especially at higher SVs. Conclusions. Stroke volume calculation from real-time 3D color Doppler data in this in vitro study shows that at depths greater than approximately 13 cm, accuracy decreases, especially at higher SVs. This may be due to decreased resolution and the reduced frame rate at these depths. At shallower depths, volume calculation form the 3D Doppler data appears very accurate.

Key Words: Doppler sonography • echocardiography • 3-dimensional

Abbreviations: ANOVA, analysis of variance • ECG, electrocardiogram • LVOT, left ventricular outflow tract • SV, stroke volume • 3DE, 3-dimensional echocardiography • 2DE, 2-dimensional echocardiography




This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
R. M. Lang, V. Mor-Avi, L. Sugeng, P. S. Nieman, and D. J. Sahn
Three-Dimensional Echocardiography: The Benefits of the Additional Dimension
J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2053 - 2069.
[Abstract] [Full Text] [PDF]