JUM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Macdonald, M. C.
Right arrow Articles by Madsen, E. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Macdonald, M. C.
Right arrow Articles by Madsen, E. L.

Journal of Ultrasound in Medicine, Vol 18, Issue 1 55-62, Copyright © 1999 by American Institute of Ultrasound in Medicine


JOURNAL ARTICLE

Acoustic measurements in a tissue mimicking liquid

M. C. Macdonald and E. L. Madsen
Acuson Corporation, Mountain View, California, USA.

A liquid has been developed that mimics soft tissue in terms of propagation speed, attenuation, and nonlinearity parameter B/A. Since the slope of the attenuation coefficient is nearly constant up to at least 18 MHz and the value can be anywhere in the range 0.1 through 0.7 dB/(cm MHz), the material is ideal for assessing the effectiveness of attenuation derating of output acoustic intensities and pressures measured in water. A clinical ultrasound system was used to transmit into the tissue mimicking liquid. The pulse intensity integral and rarefactional pressure at various field locations in the tissue mimicking liquid were computed and compared to the corresponding values measured in water with attenuation derating applied. The slope of the attenuation coefficient in the tissue mimicking liquid was used for the derating. From this data it is seen that the present practice of modeling in vivo exposure using linearly derated measurements made in water is not always conservative.


This article has been cited by other articles:


Home page
J Ultrasound MedHome page
T. A. Stiles, E. L. Madsen, G. R. Frank, T. Diehl, and J. A. Lucey
Tissue-Mimicking Liquid for Use in Exposimetry
J. Ultrasound Med., April 1, 2005; 24(4): 501 - 516.
[Abstract] [Full Text] [PDF]