Journal of Ultrasound in Medicine, Vol 16, Issue 8 529-535, Copyright © 1997 by American Institute of Ultrasound in Medicine
Prototype therapeutic ultrasound emitting catheter for accelerating thrombolysis
K. Tachibana and S. Tachibana
The aim of this study was to evaluate the performance of a new therapeutic ultrasound device for accelerating thrombolysis by urokinase. Artificial thrombi were produced in vitro by Chandler's loop method. Urokinase (240 IU) was added to the artificial thrombi (n = 5) before exposure to ultrasound with the new catheter-delivered device. Ultrasound (1.3 MHz) was applied with an intensity of 0.3 W/cm2 for 60 s. The degree of fibrinolysis was obtained by measurement of thrombus weight and calculated as percentages. Fibrinolysis was compared between control thrombus groups (n = 5) treated with urokinase alone or with ultrasound alone (analysis of variance, P < 0.05). Ultrasound plus urokinase (33.1 +/- 6.1% [mean +/- standard deviation]) was significantly different urokinase alone (12.3 +/- 2.4%) after 15 min of incubation. Ultrasound alone resulted in no fibrinolysis (0.9 +/- 2.5%). Ultrasound energy from the catheter-based device was sufficient to induce acceleration of fibrinolysis with urokinase. |
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |