JUM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Order Full text via Infotrieve
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Groth, D. S.
Right arrow Articles by Pavlicek, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Groth, D. S.
Right arrow Articles by Pavlicek, W.

Journal of Ultrasound in Medicine, Vol 14, Issue 4 273-277, Copyright © 1995 by American Institute of Ultrasound in Medicine


JOURNAL ARTICLE

Blood flow velocity measurements: a comparison of 25 clinical ultrasonographic units

D. S. Groth, F. E. Zink, J. P. Felmlee, J. M. Kofler Jr, E. M. James, J. R. Lindsey and W. Pavlicek
Department of Diagnostic Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA.

A blood-mimicking flow phantom was used to evaluate the precision of velocity measurements acquired using 25 pulsed Doppler ultrasonographic units from four vendors. Measurements were made at four constant flow rates (12 to 50 cm/s peak velocity). The average standard deviation values of the peak and time-averaged velocities among all units and all flow rates were found to be 7 and 9% of the mean, respectively, while the corresponding values for a subgroup of 20 identical units were 5 and 8%. Considered in conjunction with other published data, this suggests that units should be calibrated to an institutional standard at the time of acceptance testing.


This article has been cited by other articles:


Home page
J Ultrasound MedHome page
J. F. Soustiel, E. Levy, M. Zaaroor, R. Bibi, S. Lukaschuk, and D. Manor
A New Angle-Independent Doppler Ultrasonic Device for Assessment of Blood Flow Volume in the Extracranial Internal Carotid Artery
J. Ultrasound Med., December 1, 2002; 21(12): 1405 - 1412.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1995 by the American Institute of Ultrasound in Medicine.