Journal of Ultrasound in Medicine, Vol 12, Issue 5 285-295, Copyright © 1993 by American Institute of Ultrasound in Medicine
Sonographic heat generation in vivo in the gravid long-tailed macaque (Macaca fascicularis)
A. F. Tarantal, F. Chu, W. D. O'Brien Jr and A. G. Hendrickx
Temperature elevations that occur during diagnostic ultrasonic exposure were assessed in vivo in gravid macaques after 10, 20, or 30 min (scan mode; N = 30) or 5, 10, or 15 min (pulsed Doppler; N = 32). Five time points were assessed during the second and third trimesters (gestational days 70 to 150 +/- 2; term, approximately 165 days) using a transient thermocouple technique. Measurements were obtained intracranially or at the muscle-bone interface using a commercial sector scanner (ATL MK 600, 7.5 MHz scanhead; scan mode, ISPTA) = 27 mW/cm2, ISPPA = 85 W/cm2, pulse repetition frequency (PRF) = 1 kHz; pulsed Doppler - ISPTA = 54 mW/cm2, ISPPA = 1.5 W/cm2, PRF = 18.5 kHz). Overall, the greatest temperature elevation achieved with either modality or location was 0.6 degrees C. This article has been cited by other articles:
|
|||||||||||||||