|
|
||||||||
|
Journal of Ultrasound in Medicine, Vol 10, Issue 3 141-148, Copyright © 1991 by American Institute of Ultrasound in Medicine
Effect of tank liquid acoustic velocity on Doppler string phantom measurements
A. Goldstein
The quantitative effects of degassed water in string phantom tank Doppler measurements are derived theoretically. The Doppler parameter measurements considered are range gate registration, range gate profile, image flow angle measurements, and velocity calculation. The equipment velocity calculation is demonstrated to have an appreciable error which is due to the water acoustic velocity and the transducer acquisition geometry. A velocity calibration technique is proposed that only needs a simple multiplicative factor to compensate for the water in the tank. |
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |